Replacing System Stirrups

In this online tutorial, we demonstrate how system stirrups can be exchanged by producing a new foot piece, without having to build an entire new orthosis. For this, you can use a present cast model as your basis or produce a new one (e. g. based on the old one).

In addition, we provide further online tutorials with more information on the production of a partially flexible, rigid or slim foot piece as well as on bending the system stirrups.

As several system joints require the same tools and similar work steps, we demonstrate the respective technique for various system joints on one example.

Example NEURO SWING 2:
NEURO CLASSIC, NEURO CLASSIC-SPRING, NEURO CLASSIC-SWING, NEURO HiSWING, NEURO VARIO 2, NEURO VARIO-CLASSIC 2, NEURO VARIO-SPRING 2, NEURO VARIO-SWING, NEURO SWING, NEURO SWING 2, NEURO SWING-CLASSIC

Example NEURO VARIO- CLASSIC:
NEURO SPRING, NEURO VARIO, NEURO VARIO-CLASSIC, NEURO VARIO-SPRING

Example NEURO SWING Carbon:
NEURO SWING Carbon

You will find further information on the period of use of our system components in the maintenance plans.

Fase 1/5

In order to replace the system stirrups, you need:

  • a new system stirrup,
  • an assembly/lamination dummy corresponding to the system joint and
  • an assembly/lamination dummy for system stirrups.

Note: Mind the corresponding system width.

Fase 2/5

Remove both spacers from the assembly/lamination dummy for system stirrups.

Fase 3/5

Important! Use a new system stirrup.
First, mark the lower edge of the dummy on the system stirrup. Everything above this line must not be bent, as this area (the joint area) is a part of the system joint later.

Fase 4/5

Produce the foot piece as usual. Use the assembly/lamination dummy for system stirrups instead of the joint's upper part.

Fase 5/5

Mount the system ankle joint as described in the instructions for use and the online tutorial.

Fase 1/4

In order to replace the system stirrups, you need:

  • a new system stirrup,
  • an assembly/lamination dummy corresponding to the system joint and
  • an assembly/lamination dummy for system stirrups.

Note: Mind the corresponding system width.

Fase 2/4

Important! Use a new system stirrup.
First, mark the lower edge of the dummy on the system stirrup. Everything above this line must not be bent, as this area (the joint area) is a part of the system joint later.

Fase 3/4

Produce the foot piece as usual. Use the assembly/lamination dummy for system stirrups instead of the joint's upper part.

Fase 4/4

Mount the system ankle joint as described in the instructions for use and the online tutorial.

Fase 1/4

In order to replace the system stirrups, you need:

  • a new system stirrup and
  • a NEURO SWING Carbon assembly/lamination dummy.

Note: Mind the corresponding system width.

Fase 2/4

Important! Use a new system stirrup.
First, mark the lower edge of the dummy on the system stirrup. Everything above this line must not be bent, as this area (the joint area) is a part of the system joint later.

Fase 3/4

Produce the foot piece as usual. Work without system anchors.

Fase 4/4

Mount the system ankle joint as described in the instructions for use and the online tutorial. 

Last Update: 28 September 2022


×

Tabella cookie tecnici:

Nome Scopo Durata della memorizzazione Gestore Tipo
fe_typo_user Questo cookie è un cookie di sessione standard di TYPO3; salva i dati di accesso inseriti per il login di un utente in un’area chiusa. Durata della sessione TYPO3 Association, Sihlbruggstrasse 105, 6340 Baar, Svizzera HTTP
rc::c Distingue tra esseri umani e robot Durata della sessione Google Ireland Limited, Google Building Gordon House, 4 Barrow St, Dublin, D04 E5W5, Irland HTML
×

Tabella cookie marketing:

Nome Scopo Durata della memorizzazione Gestore Tipo
_ga Registra un ID univoco per generare dati statici per l'utilizzo del sito Internet da parte dell'utente. 2 anni Google Ireland Limited, Google Building Gordon House, 4 Barrow St, Dublino, D04 E5W5, Irlanda HTTP
_gid Registra un ID univoco per generare dati statici per l'utilizzo del sito Internet da parte dell'utente. 1 giorno Google Ireland Limited, Google Building Gordon House, 4 Barrow St, Dublino, D04 E5W5, Irlanda HTTP
_gat l cookie _gat_UA contiene inoltre il numero di identificazione (_gat_ <***>) del conto di Google Analytics o del sito Internet a cui fa riferimento il cookie. 1 giorno Google Ireland Limited, Google Building Gordon House, 4 Barrow St, Dublino, D04 E5W5, Irlanda HTTP
_gcl_au È utilizzato da Google AdSense per l’efficacia pubblicitaria sui siti Internet. 3 mesi Google Ireland Limited, Google Building Gordon House, 4 Barrow St, Dublino, D04 E5W5, Irlanda HTTP
r/collect Trasmette dati mediante il dispositivo e il comportamento dell’utente su Google Analytics. Durata della sessione Google Ireland Limited, Google Building Gordon House, 4 Barrow St, Dublino, D04 E5W5, Irlanda Pixel